Charles Explorer logo
🇨🇿

Loss of di- and trimethylarsine on Nafion membrane dryers following hydride generation;

Publikace na Přírodovědecká fakulta |
2011

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

Complete loss of trimethylarsine upon passage through a Nafion membrane dryer was identified. Pronounced loss of dimethylarsine-19 and 66%, depending on dryer dimensions-was also observed; arsine and methylarsine did not exhibit significant losses.

Arsines were generated from corresponding precursors at the 2 ng ml(-1) level by tetrahydroborate reduction, after passing through the dryer separated by cryotrapping and thermal desorption and detected by atomic absorption spectrometry with a quartz multiatomizer. Since Nafion membrane dryers have been commonly used in analytical atomic spectroscopy, implications for trace and speciation analysis of arsenic are discussed.

A dryer based on sodium hydroxide pellets is proposed as an alternative, safe for all arsines.