Charles Explorer logo
🇨🇿

In-syringe magnetic-stirring-assisted liquid-liquid microextraction for the spectrophotometric determination of Cr(VI) in waters

Publikace na Farmaceutická fakulta v Hradci Králové |
2013

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

A fully automated method for the determination of chromate is described. It is based on the selective reaction of Cr(VI) with diphenylcarbazide in acidic media to form a colored complex of Cr(III) with the oxidation product diphenylcarbazone.

The reaction was performed within the syringe of an automatic burette containing a magnetic stirrer for homogenization of the sample and the required reagents. In-syringe stirring was made possible using a specially designed driving device placed around the syringe barrel to achieve a rotating magnetic field in the syringe, forcing the stirrer to spin.

In a second step, the reaction mixture in the syringe was neutralized to allow in-syringe magnetic-stirring-assisted dispersive liquid-liquid microextraction of the complex into 125 mu L of n-hexanol. After phase separation by droplet flotation over 30 s, the organic phase was propelled into a coupled spectrophotometric detection cell.

The entire multistep procedure including in-system standard preparation was done within 270 s. The method was used for the analysis of natural waters, achieving average analyte recovery of 103 %, a limit of detection of 0.26 mu g L-1, and a repeatability of less than 4 % relative standard deviation.