Purpose: To assess the long-term mechanical stability and accuracy of the patient positioning system (PPS) of the Leksell Gamma Knife (R) Perfexion (TM) (LGK PFX). Methods: The mechanical stability of the PPS of the LGK PFX was evaluated using measurements obtained between September 2007 and June 2011.
Three methods were employed to measure the deviation of the coincidence of the radiological focus point (RFP) and the PPS calibration center point (CCP). In the first method, the onsite diode test tool with single diode detector was used together with the 4 mm collimator on a daily basis.
In the second method, a service diode test tool with three diode detectors was used biannually at the time of the routine preventive maintenance. The test performed with the service diode test tool measured the deviations for all three collimators 4, 8, and 16 mm and also for three different positions of the PPS.
The third method employed the conventional film pin-prick method. This test was performed annually for the 4 mm collimator at the time of the routine annual QA.
To estimate the effect of the patient weight on the performance of the PPS, the focus precision tests were also conducted with varying weights on the PPS using a set of lead bricks. Conclusions: Excellent long-term mechanical stability and high accuracy was observed for the PPS of the LGK PFX.
No PPS recalibration or any adjustment in the PPS was needed during the monitored period of time. Similarly, the weight on the PPS did not cause any significant disturbance in the performance of the PPS for up to 202 kg simulated patient weight.