Background: Adults with transposition of the great arteries (TGA) after atrial switch repair have an increased risk for arrhythmia and sudden cardiac death. We analyzed whether a remote monitoring (RM) system as part of an implantable cardiac device contributes to timely recognition and improved treatment of critical arrhythmias in these patients.

Methods and Results: All consecutive TGA patients (n=11) requiring a pacemaker or cardiac resynchronization therapy with or without implantable cardioverter defibrillator between 2008 and 2011 were included. AM-detected arrhythmia, abnormality of device integrity and reaction time from event transmission until acknowledgement via email and clinical decision making were analyzed and compared to a control group (n=21).

In 10 patients (91%) 17 arrhythmias were detected, 8 patients (80%) indicated no symptoms. In the RM group time interval from transmission to acknowledgement was 2.4 days (range, 0-4.5 days).

Clinical decision-making was advanced by a mean of 77.5 days (range, 10-197 days) compared with conventional follow-up and identified adaption of anti-arrhythmic medication in 8, electrical cardioversion in 2, overdrive pacing in 1 and radiofrequency ablation in 2 patients. A coronary sinus lead fracture was identified in 1 patient followed by successful replacement.