Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Meropenem Hydrolysis Assay with NH4HCO3, a Reliable Tool for Direct Detection of Carbapenemase Activity

Publikace na Lékařská fakulta v Plzni |
2015

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

A comparison of a matrix-assisted laser desorption ionization-time of flight mass spectrometric (MALDI-TOF MS) meropenem hydrolysis assay with the Carba NP test showed that both methods exhibited low sensitivity (approximately 76%), mainly due to the false-negative results obtained with OXA-48-type producers. The addition of NH4HCO3 to the reaction buffer for the MALDI-TOF MS assay dramatically improved its sensitivity (98%).

Automatic interpretation of the MALDI-TOF MS assay, using the MBT STAR-BL software, generally agreed with the results obtained after manual analysis. For the Carba NP test, spectrophotometric analysis found six additional carbapenemase producers.