Selected Ion Flow-Drift Tube Mass Spectrometry: Quantification of Volatile Compounds in Air and Breath

Publication at Faculty of Mathematics and Physics |
2015

Abstract

A selected ion flow-drift tube mass spectrometric analytical technique, SIFDT-MS, is described that extends the established selected ion flow tube mass spectrometry, SIFT-MS, by the inclusion of a static but variable E-field along the axis of the flow tube reactor in which the analytical ion molecule chemistry occurs. The ion axial speed is increased in proportion to the reduced field strength E/N (N is the carrier gas number density), and the residence/reaction time, t, which is measured by Hadamard transform multiplexing, is correspondingly reduced.

To ensure a proper understanding of the physics and ion chemistry underlying SIFDT-MS, ion diffusive loss to the walls of the flow-drift tube and the mobility of injected H3O+ ions have been studied as a function of E/N. It is seen that the derived diffusion coefficient and mobility of H3O+ ions are consistent with those previously reported.

The rate coefficient has been determined at elevated E/N for the association reaction of the H3O+ reagent ions with H2O molecules, which is the first step in the production of H3O+(H2O)(1,2,3) reagent hydrate ions. The production of hydrated analyte ion was also experimentally investigated.

The analytical performance of SIFDT-MS is demonstrated by the quantification of acetone and isoprene in exhaled breath. Finally, the essential features of SIFDT-MS and SIFT-MS are compared, notably pointing out that a much lower speed of the flow-drive pump is required for SIFDT-MS, which facilitates the development of smaller cost-effective analytical instruments for real time breath and fluid headspace analyses.