Charles Explorer logo
🇨🇿

Rapid determination of branched chain amino acids in human blood plasma by pressure-assisted capillary electrophoresis with contactless conductivity detection

Publikace na 3. lékařská fakulta |
2015

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

A CE method with contactless conductivity detection has been developed for the clinical determination of the branched chain amino acids (BCAAs) valine, isoleucine and leucine in human blood plasma. The CE separation was performed in an optimised BGE with composition of 3.2M acetic acid in 20%v/v methanol, pH2.0.

The achieved separation time was 125s when using a capillary with an effective length of 14.7cm, electric field intensity of 0.96kV/cm and simultaneous application of a hydrodynamic pressure of 50mbar. The separation efficiency in blood plasma equalled 461000 theoretical plates/m for valine and isoleucine, and 455000 theoretical plates/m for leucine; the detection limits are equal to 0.4M for all three amino acids.

The RSD values for repeatability of the migration time equalled 0.1% for measurements during a single day and 0.3% for measurements on different days; the RSD values for repeatability of the peak areas equalled 2.3-2.6% for measurements during a single day and 2.7-4.6% for measurements on different days. It followed from the performed tests that the plasmatic levels of BCAAs attain a maximum 60min after intravenous application of an infusion of BCAAs.