Charles Explorer logo
🇬🇧

Analysis of Oxidative Damage to Components Removed from Beznoska/Poldi Total Hip Replacements

Publication at First Faculty of Medicine |
2016

Abstract

PURPOSE OF THE STUDY: Ultrahigh molecular weight polyethylene (UHMWPE) is today the most frequently used bearing surface in total joint replacements (TJR) because of its properties, i.e., excellent biocompatibility, good mechanical andtribological performance and high wear resistance. UHMWPE liners are the most loaded TJR components and, therefore, their properties are decisive for TJR longevity.

This study had three objectives: 1)to evaluate the oxidative degradation of explanted UHMWPE components; 2) to look for a statistically significant relationship between the extent of oxidative degradation and the durability of joint replacements; and 3) to investigate whether the durability of a TJR was related to the type of sterilisation used in manufacture. MATERIAL AND METHODS: The study included 26 acetabular components obtained at revision arthroplasty between 2004 and 2013 from patients in whom a Beznoska/Poldi total hip replacement was used in the period from 1977 to 2002.

The average age of the patients at the time of primary implantation was 57.9 years, the average longevity of the components removed was 18.63 years (range, 6.9 to 27.9 years). RESULTS: The OI values (average and maximum) in both worn out and unworn surface areas were related to component longevity.

The difference between the oxidation index of ruptured components and that of the other components was statistically significant. Significant differences were also found between the average and maximum OI values of worn out areas and those of unworn surfaces.

The relationship between the average oxidation index in both the unworn and worn out areas of UHMWPE components and the longevity of cracked components was statistically significant. DISCUSSION: Our results show that the Ol values obtained by microspectroscopy correlated with both the microscopic damage and the longevity of UHMWPE liners and the correlation was statistically significant also in relation to the longevity of total replacements.