Charles Explorer logo
🇬🇧

Hemolysis index as a tool for determination of free hemoglobin in plasma. Hemolysis index as a tool for determination of free hemoglobin in plasma

Publication at Central Library of Charles University, Third Faculty of Medicine |
2018

Abstract

Objective: Implementation of a laboratory test of free hemoglobin in plasma into routine laboratory practice using a calculation method based on hemolysis index (HI). The test should be available 24/7 and quality control of this method should be ensured.

Materials and methods: The method was calibrated using in-house calibration standards. The patients' samples were compared to a spectrophotometric method according to Fairbanks [1] and to the HI-based method (measured on the automated analyzer Cobas 6000, Roche) using the Passing-Bablok regression.

The external quality assessment (EQA) samples (INSTAND e.V.) and the in-house internal quality control (IQC) samples, prepared just as the calibration standards, were measured in the same way. Yields relative to target (or theoretical) hemoglobin plasma concentrations were calculated for the both types of samples.

Results: The linear regression equation for calculation of free hemoglobin in plasma [free Hb (mg/l) = 11.41*HI+21.1] was obtained by calibration. This formula was utilized to compare patients' samples to the Fairbanks procedure [1] using Passing-Bablok regression (intercept 45.9 mg/L; slope 1.28).

Recovery of the Fairbanks method [1] relative to the target values of the EQA samples was 100.8 %. Recovery of the HI-based method was only 71.1%.

The HI-based method correlated with theoretical concentration of hemoglobin; the spectrophotometric method by Fairbanks [1] did not. Conclusion: The free hemoglobin calculation method utilizing HI values is an easy-to-use, low-cost and fast method that can be available to clinical departments 24/7.

The results are sufficient for timely identification of intravascular hemolysis. There are discrepant results of compared methods in patient samples, but we do not consider the bias as significant because of higher limit for organs damage (up 400 mg/l), which is caused by mechanical hemolysis.

However, the EQA samples correlate only with the Fairbanks method [1], due to different matrix properties of these samples, and therefore cannot be used for the HI-based method. Ensuring traceability between spectrophotometrically measured EQA samples and the HI-based method by means of patients' samples has been the only option for trueness verification of the implemented method so far.