Charles Explorer logo
🇨🇿

Mercury Speciation in Fish by High-Performance Liquid Chromatography (HPLC) and Post-Column Ultraviolet (UV)-Photochemical Vapor Generation (PVG): Comparison of Conventional Line-Source and High-Resolution Continuum Source (HR-CS) Atomic Absorption Spectrometry (AAS)

Publikace na Přírodovědecká fakulta |
2019

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

An ultraviolet-photochemical generator (UV-PVG) capable of post-column on-line transformation of both organic and inorganic mercury species to cold vapor (Hg-0) with subsequent detection by quartz tube-atomic absorption spectrometry (QT-AAS) was developed. Mercury(II), methylmercury(I), ethylmercury(I), and phenylmercury(I) were successfully detected after separation by reversed-phase high-performance liquid chromatography (RP-HPLC).

Two types of AAS detectors were compared. The first was a commonly used line-source instrument while the second was a high-resolution continuum source (HR-CS) AAS.

The latter provided better limits of detection: 0.47 mu g L-1 for Hg(II), 0.84 mu g L-1 for methylmercury(I), 0.80 mu g L-1 for ethylmercury(I), and 2.0 mu g L-1 for phenylmercury(I). The repeatability at 30 mu g L-1 was 3.6%, 4.1%, 6.2%, and 4.5% for these species (n = 10).

These figures of merit were comparable with those reported for more sensitive atomic fluorescence spectrometry. Nine sample extraction procedures were investigated.

Extraction by tetramethylammonium hydroxide and HCl at 75 degrees C was selected as the only method compatible with the proposed separation and detection steps providing high extraction efficiency and no changes in mercury speciation. The applicability of the proposed high-performance liquid chromatography-ultraviolet-photochemical vapor generation-quartz tube-atomic absorption spectrometry method was demonstrated using fish samples and certified reference materials (CRM) DOLT-4 (dogfish liver) and ERM-CE464 (tuna fish).

The results were comparable to those obtained by a reference method based on L-cysteine extraction and high-performance liquid chromatography-inductively coupled plasma-mass spectrometry (HPLC-ICP-MS) determination.