Background: The COMBO stent (OrbusNeich Medical, Ft. Lauderdale, Florida) is a new-generation bio-engineered drug eluting stent, combining an abluminal coating of a bioabsorbable polymer matrix for sustained release of sirolimus and luminal anti-CD34 coating for endothelial progenitor cell capture and rapid endothelialization.
Methods: The Multinational Abluminal Sirolimus Coated BiO-Engineered StenT (MASCOT) registry was a prospective post-marketing study conducted from June 2014-May 2017 across 60 centers globally. Patients were eligible if COMBO stent implantation was attempted, and they received dual antiplatelet therapy (DAPT) per local guidelines.
Follow-up was conducted by trained research staff at 1, 6 and 12 months by phone or clinic visit to capture clinical events and DAPT cessation events. The primary endpoint was 1-year target lesion failure (TLF), composite of cardiac death, non-fatal myocardial infarction not clearly attributable to a non-target vessel, or ischemia-driven target lesion revascularization.
Results: A total of 2614 patients were enrolled over the study period with 96.7% completion of 1-year follow-up. The mean age of enrolled patientswas 62.9 +/- 11.2 years and 23.0% were female.
Diabetes mellitus was present at baseline in 33.5%. A total of 56.1% patients underwent PCI for acute coronary syndrome (ACS).
The 1-year primary endpoint of TLF occurred in 3.4% patients (n = 88). Definite stent thrombosis occurred in 0.5% patients (n= 12).
Conclusion: The MASCOT post marketing registry provides comprehensive safety and efficacy outcomes following contemporary PCI using the novel COMBO stent in an all-comer population. This platform is associated with low rates of 1-year TLF and ST.
Clinicaltrials.gov identifier: NCT02183454.