Rapid and reliable HILIC-MS/MS method for monitoring allantoin as a biomarker of oxidative stress

Publikace na Přírodovědecká fakulta, 1. lékařská fakulta |
2020

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

Allantoin is an excellent biomarker of oxidative stress in humans as the main product of uric acid oxidation by reactive oxygen species. Yet, allantoin determination is still not routinely performed in clinical laboratories.

Therefore, we developed a fast, simple, selective, and sensitive UHPLC-MS/MS method for allantoin determination in human serum using an isotopically labeled internal standard. Our analytical protocol provided high sensitivity by mass spectrometry detection and high throughput by HILIC-MS/MS analysis within 4 min, with one-step serum sample preparation approximately within 7 min.

Lastly, our protocol was fully validated to demonstrate its reliability in allantoin determination in human serum. The method showed an excellent linear range from 0.05 to 100 mu M, with precision ranging from 1.8 to 11.3% (RSD), and with accuracy (relative error %) within +/- 6.0%.

The method was then applied to analyze the concentration of allantoin in serum samples from 71 patients with chronic gout without treatment with xanthine oxidase inhibitors. The median serum allantoin concentration in the cohort was 2.8 mu M (n = 71).

Overall, our simple analytical protocol has the potential to be easily implemented in clinical routine practice for monitoring allantoin as a key oxidative stress biomarker.