Charles Explorer logo
🇨🇿

The use of non-traditional carbon film electrode based on microcrystalline natural graphite - polystyrene composite film for amperometric determination of 5-aminoquinoline using flow injection analysis minimising electrode fouling

Publikace na Přírodovědecká fakulta |
2021

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

Flow injection analysis (FIA) with amperometric detection (AD) at a novel type of carbon film electrode (CFE) based on microcrystalline natural graphite ? polystyrene composite film was used for fast and sensitive determination of genotoxic 5-aminoquinoline. CFE is a recently developed electrode with large potential window, good sensitivity, and easily renewable surface.

Under optimized conditions, limit of quantitation (LOQ) of 1?10?6 mol L?1 and RSD of 2.48% was achieved and problems with working electrode passivation were successfully minimised.