Background: Left atrial appendage (LAA) closure (LAAC) is accompanied by a high risk of complications. Due to the complex anatomy of the LAA and the oval-shaped ostium, the proper sizing of the device is often difficult.
Purpose: To assess individualized fluoroscopy viewing angles using pre-procedural CT analysis and to compare the results of landing zone measurements obtained from CT, transesophageal echocardiography (TEE), intracardiac echocardiography (ICE), and fluoroscopy. Methods: Patients with indications for LAAC were enrolled.
Cardiac CT and TEE were done before the procedure; ICE and fluoroscopy measurements were done peri-procedurally. Multiplanar reconstruction of CT images, using FluoroCT software, was done, and optimal "personalized" viewing angles for fluoroscopy were determined.
Moreover, a mean (using multiplanar CT reconstruction, derived from the LAA perimetr) amd maximum (using all four imaging modalitities) landing zone (LZ) of the LAA were masured. Results: Twenty-five patients were analyzed.
Despite significant correlation between LZs obtained from different imaging modalities, the values of LZs differed significantly; the mean LZ diameter on CT was 20.60 +- 3.42 mm, the maximum diameters were 21.99 +- 4.03 mm (CT), 18.72 +- 2.44 mm (TEE), 18.20 +- 2.68 mm (ICE), and 17.76 +- 3.24 mm (fluoroscopy). The mean CT diameter matched with the final device selection in 92% patients, while fluoroscopy or TEE maximum diameters in only 72% patients.
Optimal viewing angles differed significantly from the fluoroscopy projections usually recommended by the manufacturer in 3 patients. Conclusions: CT provides the best measurement of the LZ and the best prediction of the optimum fluoroscopy projections for the implantation procedure.