Charles Explorer logo
🇬🇧

Effects of Acute Pump Speed Changes on Cerebral Hemodynamics in Patients With an Implantable Continuous-Flow Left Ventricular Assist Devices

Publication at Faculty of Physical Education and Sport, First Faculty of Medicine, Second Faculty of Medicine, Third Faculty of Medicine |
2021

Abstract

Mechanical circulatory support (MCS) with an implantable left ventricular assist device (LVAD) is an established therapeutic option for advanced heart failure. Most of the currently used LVADs generate a continuous stream of blood that decreases arterial pulse pressure.

This study investigated whether a change of the pulse pressure during different pump speed settings would affect cerebral autoregulation and thereby affect cerebral blood flow (CBF). The study included 21 haemodynamically stable outpatients with a continuous-flow LVAD (HeartMate II, Abbott, USA) implanted a median of 6 months before the study (interquartile range 3 to 14 months).

Arterial blood pressure (measured by finger plethysmography) was recorded simultaneously with CBF (measured by transcranial Doppler ultrasound) during baseline pump speed (8900 rpm [IQR 8800; 9200]) and during minimum and maximum tolerated pump speeds (8000 rpm [IQR 8000; 8200] and 9800 rpm [IQR 9800; 10 000]). An increase in LVAD pump speed by 800 rpm [IQR 800; 1000] from the baseline lead to a significant decrease in arterial pulse pressure and cerebral blood flow pulsatility (relative change -24 % and -32 %, both p < 0.01), but it did not affect mean arterial pressure and mean CBF velocity (relative change 1 % and -1.7 %, p=0.1 and 0.7).

In stable patients with a continuous-flow LVAD, changes of pump speed settings within a clinically used range did not impair static cerebral autoregulation and cerebral blood flow.