Total body irradiation for standard treatment rooms: a robust sweeping beam technique with respect to the body shape

Publikace |
2022

Tento text není v aktuálním jazyce dostupný. Zobrazuje se verze "en".Abstrakt

The purpose of this work is to improve a sweeping beam technique for total body irradiation (TBI) on a low flat couch using a varying patient thickness model. We designed a flat couch for total body irradiation in supine and prone position Three generic arcs with rectangular segments for a patient torso thickness of 16, 22 and 28 cm were generated with respect to varying patient thickness of four particular parts of the body: head, torso, thighs and calves.

Measured dose homogeneity stays within +-10% of prescribed dose for all of the three patient thickness models. The robustness of the method was evaluated as the increase in dose in the phantom center of 0.7% per 1 cm reduction in phantom thickness.