In patients with heart failure and preserved or mildly reduced ejection fractions (EF >= 40%), implantation of an interatrial shunt device (IASD) resulted in heterogenous changes of the left atrial (LA) volume. Baseline characteristics that correlate with a favorable decrease in LA volume are unknown.
We hypothesized that a larger ratio of left to right atrial volume at baseline would correlate strongly with LA volume decongestion following IASD implantation. Reduce Elevated LA Pressure in Patients With Heart Failure was a multicenter study of the safety and feasibility of IASD implantation.
Sixty-four patients with EF >= 40% underwent device implantation along with baseline conventional echocardiograms, speckle tracking echocardiography, and resting and exercise hemodynamics. Higher LA compliance (-4.2%, p = 0.048) and right atrial reservoir strain (-0.8%, p = 0.005) were independently associated with a percent decrease in the systolic LA volume index from baseline to 6-months.
In conclusion, greater LA volume reduction following IASD implantation is associated with higher baseline compliance of the left atrium and higher reservoir strain of the right atrium.